Notice ID
903202a3b7514bf5a8c3a14d2713ea25
This solicitation is for a XYZ three axis manipulator to position a xray detector in a vacuum chamber 1E7 mbar The device will bolt to a fixed flange on the side of the vacuum chamber and the moving flange to an xray detector The following specifications are to be met X and Y travel of 25 mm with 5 micron or better resolution Z travel of 400 mm with 5 micron or better resolution Clear aperture shall allow for passage of a 38 mm diameter sensor head The fixed flange is a 6 OD CF and the moving flange is 45 OD CF All axes are stepper motor driven All axes must have absolute encoding All axes must be guided by profile rail type guides no ball bushings are allowed The moving flange load capacity should be 137 N in the vertical and 14 Nm about the horizontal axis
-
Department/Ind.Agency
ENERGY, DEPARTMENT OF
-
Subtier
ENERGY, DEPARTMENT OF
-
Office
ARGONNE NATL LAB - DOE CONTRACTOR